Combined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study

Abstract: BackgroundCombined effects of leaks, mechanical property of respiratory system and upper airway (UA) patency on patient-ventilator synchrony (PVA) and the level of clinically “tolerable” leaks are not well established in home ventilators.MethodsWe comparatively assessed on a bench model, the highest leak level tolerated without inducing significant asynchrony (“critical leak”) in three home ventilators (Astral 150, Trilogy 100 and Vivo 60; noted as A150, T100 and V60 respectively) subjected to three simulated … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
22
0
3

Year Published

2018
2018
2024
2024

Publication Types

Select...
5
3

Relationship

0
8

Authors

Journals

citations
Cited by 25 publications
(30 citation statements)
references
References 32 publications
0
22
0
3
Order By: Relevance
“…studies investigating NIV leakage in a pediatric population; furthermore, the level of clinical "tolerability" of leaks is not well established in home NIV. 40 Despite this, it is widely known that large degrees of leakage will significantly affect NIV performance, as well as comfort. 41 Large air leaks can significantly reduce the level of delivered intra-alveolar pressure, indeed the response of the PAP machine can be reduced by up to 56% in the presence of a leak of 30 L/min.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…studies investigating NIV leakage in a pediatric population; furthermore, the level of clinical "tolerability" of leaks is not well established in home NIV. 40 Despite this, it is widely known that large degrees of leakage will significantly affect NIV performance, as well as comfort. 41 Large air leaks can significantly reduce the level of delivered intra-alveolar pressure, indeed the response of the PAP machine can be reduced by up to 56% in the presence of a leak of 30 L/min.…”
Section: Discussionmentioning
confidence: 99%
“…Having established satisfactory compliance within our DS pediatric cohort, we next looked at another key factor of therapeutic success, effective NIV delivery, and system leakage. To date, there have been no studies investigating NIV leakage in a pediatric population; furthermore, the level of clinical “tolerability” of leaks is not well established in home NIV 40 . Despite this, it is widely known that large degrees of leakage will significantly affect NIV performance, as well as comfort 41 .…”
Section: Discussionmentioning
confidence: 99%
“…With the proliferation of various NIV devices intended for both hospital and home use over the last 2 decades, much interest has focused on improving patient-ventilator interaction. 8,[18][19][20][21] Patient-ventilator asynchrony is common in NIV, with some studies suggesting a clinical prevalence of 25-50%. 8,22 In one study, 12% of breaths delivered to 53 subjects receiving invasive mechanical ventilation (for a variety of indications) showed evidence of double-triggering.…”
Section: Discussionmentioning
confidence: 99%
“…No esta tan claro el límite para hablar de una fuga exageradamente alta, se ha reportado una fuga intencional programada de aproximadamente 200ml/segundo (12 L/min) a PEEP 5. Los ventiladores Trilogy®, de acuerdo al fabricante, son capaces de compensar hasta el doble de su fuga, una fuga de hasta aproxima-damente 25 L/min podría ser tolerada, siempre considerando que debe chequearse el motivo de la fuga(12)(19) (21)(22)(23).…”
Section: Conclusionesunclassified