2009
DOI: 10.1097/ana.0b013e31818f2eac
|View full text |Cite
|
Sign up to set email alerts
|

Clinical Evaluation of a New Multiparameter Neuromonitoring Device: Measurement of Brain Tissue Oxygen, Brain Temperature, and Intracranial Pressure

Abstract: Combining 3 different neuromonitoring functions in 1 probe might ease monitoring by making a second (ptiO2) probe unnecessary. Interpretation of our data is limited by several factors: (1) monocentric study; (2) reduced mechanical probe stability, handling difficulties with the double lumen bolt; (3) design changes to improve mechanical stability will require further study; (4) conflict of interest with Raumedic because of its support for the study. The conclusion drawn from our study is that the new multipara… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
36
0

Year Published

2011
2011
2021
2021

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 33 publications
(37 citation statements)
references
References 9 publications
0
36
0
Order By: Relevance
“…A one-on-one direct intraparenchymal comparison between Licox and Neurovent in 33 patients showed no relevant differences between the Licox (19.5±7.1 mmHg) and Neurovent multiparameter probe derived values (21.7±9.5 mmHg) [71]. This was not confirmed in a preliminary report in 11 mixed TBI and SAH patients [261].…”
Section: Definition Of Normal and Pathological Values -A Complex Taskmentioning
confidence: 76%
See 3 more Smart Citations
“…A one-on-one direct intraparenchymal comparison between Licox and Neurovent in 33 patients showed no relevant differences between the Licox (19.5±7.1 mmHg) and Neurovent multiparameter probe derived values (21.7±9.5 mmHg) [71]. This was not confirmed in a preliminary report in 11 mixed TBI and SAH patients [261].…”
Section: Definition Of Normal and Pathological Values -A Complex Taskmentioning
confidence: 76%
“…It has also been used in patients with spontaneous intraparenchymal hemorrhage [69][70][71][72][73][74][75][76], ischemic stroke in children and adults [36,71,77], severe haemorrhagic stroke [78], and in silent infarcts after subarachnoid hemorrhage [79].…”
Section: Pbto2mentioning
confidence: 99%
See 2 more Smart Citations
“…Again, with the Neurovent-PTO, higher P btO 2 values were noted, especially when testing in high P aO 2 situations. 27,40,41 Differences between the probes exist and have been partially attributed to the different sampling sizes of the probes. 40,41 The Licox has a sampling size of 13 mm 2 , whereas the Neurovent-PTO samples 22 mm 2 of the surrounding brain tissue.…”
Section: The Technology Of the Partial Pressure Of Brain Oxygenmentioning
confidence: 99%