Abstract:Background:
Contactless photoplethysmography (PPG) potentially affords the ability to obtain vital signs in pediatric populations without disturbing the child. Most validity studies have been conducted in laboratory settings or with healthy adult volunteers. This review aims to evaluate the current literature on contactless vital signs monitoring in pediatric populations and within a clinical setting.
Methods:
OVID, Webofscience, Cochrane library, and clinicaltrials.org were systematical… Show more
“…To our best knowledge, this is the first study to evaluate the use of rPPG in obtaining paediatric vital signs across varying age groups. Bautista et al published a systematic review and meta-analysis of rPPG in children in May 2023 ( 29 ). 14 of the 15 studies included in the review were performed solely in the neonatal population.…”
Background
Advancements in medical technologies have led to the development of contact-free methods of haemodynamic monitoring such as remote photoplethysmography (rPPG). rPPG uses video cameras to interpret variations in skin colour related to blood flow, which are analysed to generate vital signs readings. rPPG potentially ameliorates problems like fretfulness and fragile skin contact associated with conventional probes in children. While rPPG has been validated in adults, no prior validation has been performed in children.
Methods
A two-phased prospective cross-sectional single-centre study was conducted from January to April 2023 to evaluate the feasibility, acceptability, and accuracy of obtaining heart rate (HR), respiratory rate (RR) and oxygen saturation (SpO
2
) using rPPG in children, compared to the current standard of care. In Phase 1, we recruited patients ≤16 years from the neonatal and paediatric wards. We excluded preterm neonates with gestational age <35 weeks and newborns <24 hours old. The rPPG webcam was positioned 30 cm from the face. After 1 minute of facial scanning, readings generated were compared with pulse oximetry for HR and SpO
2
, and manual counting for RR. Correlation and Bland-Altman analyses were performed. In Phase 2, we focused on the population in whom there was potential correlation between rPPG and the actual vital signs.
Results
Ten neonates and 28 children aged 5 to 16 years were recruited for Phase 1 (765 datapoints). All patients were haemodynamically stable and normothermic. Patients and caregivers showed high acceptability to rPPG. rPPG values were clinically discrepant for children <10 years. For those ≥10 years, moderate correlation was observed for HR, with Spearman’s correlation coefficient (Rs) of 0.50 [95% confidence intervals (CI): 0.42, 0.57]. We performed Phase 2 on 23 patients aged 12 to 16 years (559 datapoints). Strong correlation was observed for HR with Rs=0.82 (95% CI: 0.78, 0.85). There was weak correlation for SpO
2
and RR (Rs=−0.25 and −0.02, respectively).
Conclusions
Our study showed that rPPG is acceptable and feasible for neonates and children aged 5 to 16 years, and HR values in older children aged 12 to 16 years correlated well with the current standard. The rPPG algorithms need to be further refined for younger children, and for obtaining RR and SpO
2
in all children. If successful, rPPG will provide a viable contact-free alternative for assessing paediatric vital signs, with potential use in remote monitoring and telemedicine.
“…To our best knowledge, this is the first study to evaluate the use of rPPG in obtaining paediatric vital signs across varying age groups. Bautista et al published a systematic review and meta-analysis of rPPG in children in May 2023 ( 29 ). 14 of the 15 studies included in the review were performed solely in the neonatal population.…”
Background
Advancements in medical technologies have led to the development of contact-free methods of haemodynamic monitoring such as remote photoplethysmography (rPPG). rPPG uses video cameras to interpret variations in skin colour related to blood flow, which are analysed to generate vital signs readings. rPPG potentially ameliorates problems like fretfulness and fragile skin contact associated with conventional probes in children. While rPPG has been validated in adults, no prior validation has been performed in children.
Methods
A two-phased prospective cross-sectional single-centre study was conducted from January to April 2023 to evaluate the feasibility, acceptability, and accuracy of obtaining heart rate (HR), respiratory rate (RR) and oxygen saturation (SpO
2
) using rPPG in children, compared to the current standard of care. In Phase 1, we recruited patients ≤16 years from the neonatal and paediatric wards. We excluded preterm neonates with gestational age <35 weeks and newborns <24 hours old. The rPPG webcam was positioned 30 cm from the face. After 1 minute of facial scanning, readings generated were compared with pulse oximetry for HR and SpO
2
, and manual counting for RR. Correlation and Bland-Altman analyses were performed. In Phase 2, we focused on the population in whom there was potential correlation between rPPG and the actual vital signs.
Results
Ten neonates and 28 children aged 5 to 16 years were recruited for Phase 1 (765 datapoints). All patients were haemodynamically stable and normothermic. Patients and caregivers showed high acceptability to rPPG. rPPG values were clinically discrepant for children <10 years. For those ≥10 years, moderate correlation was observed for HR, with Spearman’s correlation coefficient (Rs) of 0.50 [95% confidence intervals (CI): 0.42, 0.57]. We performed Phase 2 on 23 patients aged 12 to 16 years (559 datapoints). Strong correlation was observed for HR with Rs=0.82 (95% CI: 0.78, 0.85). There was weak correlation for SpO
2
and RR (Rs=−0.25 and −0.02, respectively).
Conclusions
Our study showed that rPPG is acceptable and feasible for neonates and children aged 5 to 16 years, and HR values in older children aged 12 to 16 years correlated well with the current standard. The rPPG algorithms need to be further refined for younger children, and for obtaining RR and SpO
2
in all children. If successful, rPPG will provide a viable contact-free alternative for assessing paediatric vital signs, with potential use in remote monitoring and telemedicine.
“…Opportunities for DHTs to generate RWD/RWE [15,[21][22][23] Digital health technologies (DHTs) offer a host of opportunities to change how healthcare is delivered, tracked, and disseminated. DHTs are defined by FDA as integrated systems that use computing platforms, connectivity, software, and/or sensors for healthcare.…”
Section: Rwd/rwe To Advance Health Outcomes and Address Health Dispar...mentioning
confidence: 99%
“…Yet there are many examples to point to as to how the field is maturing. For example, in this issue, Bautista et al discuss the state of photoplethysmography, an optical technique that enables measurement of parameters associated with cardiorespiratory function [ 21 , 23 ]. They highlight the unique nature of this tool and the methods (contact vs contactless) used currently both in terms of their positive and negative attributes for specific use cases ( e.g., validation of measurement) and populations ( e.g., neonates vs adults).…”
Section: Opportunities For Dhts To Generate Rwd/rwe [
15
...mentioning
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.