2002
DOI: 10.1097/00062752-200209000-00004
|View full text |Cite
|
Sign up to set email alerts
|

Clinical application of the PFA-100®

Abstract: The PFA-100 (platelet function analyzer) is a relatively new tool for the investigation of primary hemostasis. Recent studies have shown its utility as a screening tool for investigating possible von Willebrand disorder (VWD) and various platelet disorders. More recently, the PFA-100 has been shown to be valuable in monitoring desmopressin acetate (DDAVP) therapy in both VWD and platelet disorders. The PFA-100 has also been evaluated in many other studies for its utility in assessing drug effects, for potentia… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

6
132
0
4

Year Published

2005
2005
2017
2017

Publication Types

Select...
5
4

Relationship

0
9

Authors

Journals

citations
Cited by 179 publications
(142 citation statements)
references
References 83 publications
6
132
0
4
Order By: Relevance
“…The PFA-100 is a microprocessor-controlled instrument/test cartridge system used to assess platelet function simulating platelet-based primary hemostasis in vitro (15). A syringe aspirates citrated whole blood under high shear-flow conditions (5,000 -6,000 s Ϫ1 ) through a small aperture (150 m) cut into a membrane placed in the test cartridge.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The PFA-100 is a microprocessor-controlled instrument/test cartridge system used to assess platelet function simulating platelet-based primary hemostasis in vitro (15). A syringe aspirates citrated whole blood under high shear-flow conditions (5,000 -6,000 s Ϫ1 ) through a small aperture (150 m) cut into a membrane placed in the test cartridge.…”
Section: Methodsmentioning
confidence: 99%
“…After 300 s the process automatically terminates. A CEPI-CT Յ193 s (upper normal range limit) has been considered as a cut-off value to define low responders to aspirin and was used accordingly in the present study (15,16 …”
Section: Methodsmentioning
confidence: 99%
“…aspirin) and purinergic receptor inhibitors (e.g. clopidogrel) are being used primarily for arterial disorders which are characterized by high shear rate [21,22]. However, there are many preanalytical factors such as platelet function, platelet count, red blood cells and plasma von Willebrand factor are known to affect the PFA-100 results.…”
Section: Discussionmentioning
confidence: 99%
“…The time taken to occlude the hole is the closure time (CT), a measure of overall platelet-related haemostasis and this interval will be prolonged depending on the platelet activity. The use of two different cartridges with distinct agonists allows to distinguish the platelet function alterations due to intrinsic defects (principally by using CADP cartridge) or to antiplatelet therapy with ASA (CEPI cartridge) [93][94][95][96][97] , whereas the new Innovance cartridge is affected by thienopyridine therapy [98] . In comparison to BT test, this method is revealed more sensitive [27,46] especially for diagnosis of VWD and platelet function defects [25] .…”
Section: Platelet Function Point-of-care Testingmentioning
confidence: 99%