DOI: 10.1088/1741-2552/aaa569
|View full text |Cite
|
Sign up to set email alerts
|

Chronic implantation of cuff electrodes on the pelvic nerve in rats is well tolerated and does not compromise afferent or efferent fibre functionality

Abstract: Chronic implantation of cuff electrodes on the pelvic nerve was found to be a well-tolerated procedure in rats and high frequency stimulation did not lead to loss of nerve functionality. Pelvic nerve stimulation has development potential for normalizing voiding dysfunction in conscious rats.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
6
0

Year Published

2018
2018
2020
2020

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(6 citation statements)
references
References 22 publications
0
6
0
Order By: Relevance
“…Cuff electrodes were fabricated in-house from platinum iridium wire with a cobalt core embedded in silicone epoxy (Med-4213, Nusil, Carpinteria, United States) to the following dimensions: ID 0.5 mm, OD 1.2 mm, cuff length: 2 mm, inter-electrode distance: 1 mm, lead length: 200 mm ( Crook et al, 2018 ). Bladder catheters were made from polyethylene tubing (ID = 0.76 mm; OD = 1.22 mm; length: 300 mm) with a “stop collar” at one end made by tying 6.0 suture thread round the tubing 5 mm from the tip.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Cuff electrodes were fabricated in-house from platinum iridium wire with a cobalt core embedded in silicone epoxy (Med-4213, Nusil, Carpinteria, United States) to the following dimensions: ID 0.5 mm, OD 1.2 mm, cuff length: 2 mm, inter-electrode distance: 1 mm, lead length: 200 mm ( Crook et al, 2018 ). Bladder catheters were made from polyethylene tubing (ID = 0.76 mm; OD = 1.22 mm; length: 300 mm) with a “stop collar” at one end made by tying 6.0 suture thread round the tubing 5 mm from the tip.…”
Section: Methodsmentioning
confidence: 99%
“…A length of the postganglionic nerve bundle was carefully separated from the wall of the uterus for about 2 mm proximal to the pelvic ganglion without disturbing the nearby nerves that emanate from the ganglion ( Figure 1 ). A 6.0 silk suture thread was looped under the nerve and the cuff electrode slipped under the nerve and closed to achieve a snug, but not tight, fit by tying the suture thread round the nerve cuff ( Crook et al, 2018 ). A drop of dental silicone (Klasse 4) was used to cover the nerve-cuff assembly.…”
Section: Methodsmentioning
confidence: 99%
“…Preclinical studies in animal models suggest that the pelvic nerve (rodent homolog of human pelvic splanchnic nerves) is a potential neuromodulation target for neuro-urological and other pelvic functional disorders (Crook and Lovick, 2017;Langdale et al, 2017;Crook et al, 2018;Langdale et al, 2020). For example, in cats and rodents, these paired nerves contain the majority of external sacral sensory and autonomic motor projections needed for bladder sensation and contraction.…”
Section: Introductionmentioning
confidence: 99%
“…We acknowledge that the success rate of our experiments can be improved with further optimization of the attachment of the electrode to the nerve. Although there is published evidence indicating that chronic cuff electrodes on pelvic nerve were functionally stable for electrical stimulation for several days post-surgery 44 , it remains unclear if extraneural interfaces can maintain the low impedances required for chronic neural recording. Notably, peripheral nerve recording for predicting bladder pressure remains far from clinical application due to technical and surgical challenges, especially with regards to the electrode-nerve interface as well as the lack of understanding between peripheral nerve signals and bladder volume in human subjects.…”
Section: Discussionmentioning