Characterization of positional accuracy of a double‐focused and double‐stack multileaf collimator on an MR‐guided radiotherapy (MRgRT) Linac using an IC‐profiler array

Abstract: Purpose With advance magnetic resonance (MR)‐guided online adaptive radiotherapy (MRgoART) relying on calculation‐based intensity‐modulated radiation therapy (IMRT) quality assurance (QA), accurate and sensitive QA of the multileaf collimator (MLC) becomes an increasingly essential component for routine machine QA. As such, it is important to assure compliance with the AAPM TG142 guidelines to supplement calculation‐based QA methods for an online adaptive radiotherapy program. We have developed and implemented… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
28
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
6

Relationship

3
3

Authors

Journals

citations
Cited by 20 publications
(33 citation statements)
references
References 31 publications
(73 reference statements)
1
28
0
Order By: Relevance
“…Our MLC positional accuracy is in agreeance to our institutional weekly QA MLC positional reproducibility . Small MLC gravitational‐induced effects combined with isocenter dependency with gantry rotation are observed, and are amplified due to the high sensitivity of the small‐field (i.e., 0.40 cm 2 x 0.83 cm 2 ) daily QA procedure.…”
Section: Discussionsupporting
confidence: 85%
See 3 more Smart Citations
“…Our MLC positional accuracy is in agreeance to our institutional weekly QA MLC positional reproducibility . Small MLC gravitational‐induced effects combined with isocenter dependency with gantry rotation are observed, and are amplified due to the high sensitivity of the small‐field (i.e., 0.40 cm 2 x 0.83 cm 2 ) daily QA procedure.…”
Section: Discussionsupporting
confidence: 85%
“…MLC positional accuracy at central axis and off axis was previously reported at our institution by Mittauer et al . using an ionization profiler and found to be within 0.06 ± 0.16 mm at baseline with no drift 0.00 ± 0.12 mm (n = 6 months) quantified using half beam block at cardinal angles about the central axis …”
Section: Methodsmentioning
confidence: 94%
See 2 more Smart Citations
“…However, for multileaf collimator (MLC) position verification, techniques have demonstrated high spatial sensitivity using an ion chamber array. 8 Plastic scintillators have been shown to be less affected than ion chambers by the presence of a transverse magnetic field 9 and therefore may be a strong candidate for MR-linac dosimetry. 1 Radiochromic film provides an MR-compatible high-resolution solution for many routine QA checks, however films are one-time use, must be replaced for different tests, and require post exposure readout procedures.…”
Section: Introductionmentioning
confidence: 99%