2022
DOI: 10.1088/1742-6596/2165/1/012042
|View full text |Cite
|
Sign up to set email alerts
|

Characterization of manufacturing indicator electrodes coated PVA-enzyme coated 1X GA and 3X GA 2.9 % coated PVC-KTpClPB coated using potentiometry method

Abstract: Research has been carried out on the manufacture of tungsten indicator electrodes with a diameter of 1 mm coated with PVA-Enzyme coated with PVC-KTpClPB using the biosensor potentiometer method. It has a sensitivity of 19,096 mV/decade, a detection range of 1.10-5 - 5.10-4 M, a detection limit of 1.10-5 M and a correlation coefficient of R2 = 0.9431. This detection range is very small, only a difference of one order to the negative power. On the basis of this small range, the PVA-Enzyme and PVA-Enzyme-GA solut… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(2 citation statements)
references
References 15 publications
0
2
0
Order By: Relevance
“…According to (Hakim, 2022) FTIR analysis of PVA-Enzyme/GA/PVC-KTpClPB-o-NPOE membrane, the best composition was o-NPOE 61%, and XRD diffraction spectrum pattern analysis showed a change in activity.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…According to (Hakim, 2022) FTIR analysis of PVA-Enzyme/GA/PVC-KTpClPB-o-NPOE membrane, the best composition was o-NPOE 61%, and XRD diffraction spectrum pattern analysis showed a change in activity.…”
Section: Introductionmentioning
confidence: 99%
“…The best composition of the enzyme urease 6 mg 10 mL (50% water: 50% alcohol), 35 mg PVA, 50 mg, 35 mg PVC, 50 mg KTpClPB, plasticizer (o-NPOE) 61% (Hakim, 2022). The following is the analysis of the PVA-Enzyme/GA/PVC-KTpClPB-o-NPOE indicator electrodes.…”
Section: Introductionmentioning