Characterization of a 2 × 2 array of large square bars of LaBr3:Ce detectors with γ-rays up to 22.5 MeV

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 10 publications
(2 citation statements)
references
References 37 publications
0
2
0
Order By: Relevance
“…The energy resolution of a 2″ × 2″ × 8″ LaBr 3 (Ce) crystal has been measured by Dhibar et al . 56 at several photon energies up to 4.433 MeV. Above ~2 MeV the energy resolution is around 3% FWHM (Full Width Half Maximum).…”
Section: Methodsmentioning
confidence: 98%
“…The energy resolution of a 2″ × 2″ × 8″ LaBr 3 (Ce) crystal has been measured by Dhibar et al . 56 at several photon energies up to 4.433 MeV. Above ~2 MeV the energy resolution is around 3% FWHM (Full Width Half Maximum).…”
Section: Methodsmentioning
confidence: 98%
“…Uncertainty also originates from the photo peak detection efficiency, estimated using GEANT4 simulations. We have worked extensively in characterisation and simulations of various inorganic scintillators over a wide range of energies [34,38,39]. The photo peak detection efficiency uncertainty was determined to be ∼1%.…”
Section: Protonmentioning