2014
DOI: 10.1615/jlongtermeffmedimplants.2014011133
|View full text |Cite
|
Sign up to set email alerts
|

Catastrophic Stem Taper Wear in Ceramic on Polyethylene Bearing Couple: A Case Report

Abstract: A severe abrasion occurred at the stem taper 13 years after total hip arthroplasty. The bearing couple was 28 mm ceramic on polyethylene. The ceramic head had almost no damage, whereas the stem taper had severe damage. An ectopic bone probably contributed to the abrasion. The ectopic bone had grown to restrict the normal motion in the bearing couple and all stress and torque seemed to concentrate in the junction until the junction obtained mobility instead of the original bearing couple.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(1 citation statement)
references
References 0 publications
0
1
0
Order By: Relevance
“…Only a few trunnion failures coupled with ceramic heads have been reported in the literature, while nearly all reported trunnionosis cases occurred with CoCr alloy heads . Retrieval studies and in vitro testing of ceramic heads suggest a reduction, but not the elimination, of trunnion fretting and corrosion .…”
Section: Discussionmentioning
confidence: 99%
“…Only a few trunnion failures coupled with ceramic heads have been reported in the literature, while nearly all reported trunnionosis cases occurred with CoCr alloy heads . Retrieval studies and in vitro testing of ceramic heads suggest a reduction, but not the elimination, of trunnion fretting and corrosion .…”
Section: Discussionmentioning
confidence: 99%