Results: Intra-and inter-day imprecision of the Fujirebio 25(OH)D assay was < 5%. Fujirebio 25(OH)D assay showed the highest correlation among the assays tested with the LC-MS/MS method (R = 0.986). The mean relative bias obtained was -15.6% (Fujirebio), -12.7% (Beckman), -2.1% (Abbott) and 9.7% (Roche) as compared to LC-MS/ MS. Comparison with the Labquality certified reference serum panel yielded a mean bias of -11.8% (Fujirebio), -14.1% (Beckman), 4.4% (Abbott) and 3.2% (Roche), respectively. Compared to LC-MS/MS, the sensitivity of different methods in detecting vitamin D insufficiency ( < 50 nmol/L) varied from 100% for the Fujirebio assay to 72.7% for Roche, and specificity ranged from 94.4% for Roche to 87.6% for Beckman.