2009
DOI: 10.1016/j.jvir.2009.06.022
|View full text |Cite
|
Sign up to set email alerts
|

Balloon-assisted Removal of Tilted Inferior Vena Cava Filters with Embedded Tips

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
27
0

Year Published

2011
2011
2023
2023

Publication Types

Select...
4
4

Relationship

0
8

Authors

Journals

citations
Cited by 46 publications
(30 citation statements)
references
References 9 publications
0
27
0
Order By: Relevance
“…Wire loop-and-snare Tilted or embedded filters can often be retrieved using the guidewire loop and snare technique described by Rubenstein (16) and others (17). A reverse-curve catheter is placed in IVC below the level of the filter and is used to direct a guidewire through filter legs so that the guidewire tip courses cephalad underneath the filter apex where the legs join.…”
Section: Single Access Techniquesmentioning
confidence: 99%
See 2 more Smart Citations
“…Wire loop-and-snare Tilted or embedded filters can often be retrieved using the guidewire loop and snare technique described by Rubenstein (16) and others (17). A reverse-curve catheter is placed in IVC below the level of the filter and is used to direct a guidewire through filter legs so that the guidewire tip courses cephalad underneath the filter apex where the legs join.…”
Section: Single Access Techniquesmentioning
confidence: 99%
“…An angioplasty balloon is then advanced over the guidewire between the embedded filter and the caval wall and is inflated in order to realign the filter more centrally within the IVC. This may then provide a better angle for snaring the filter hook (17,18). Another advantage of this technique is that even partial tracts that are created by the guidewire may be used to sequentially separate the embedded areas of the IV filter from the caval wall in a piecemeal fashion.…”
Section: Balloon-displacement Techniquementioning
confidence: 99%
See 1 more Smart Citation
“…A variety of factors are known be associated with higher routine retrieval failure rates including longer dwell time, more transverse tilt, and presence of an embedded hook (1)(2)(3)(4). It has been postulated that fibrin caps form around the IVC filter apex when their position is near the caval wall (5,6). This position creates local eddy currents, which promote the deposition of fibrin, similar to the fibrin sheaths that form around indwelling venous catheters.…”
Section: Discussionmentioning
confidence: 99%
“…The balloon displacement technique can be used with or without a through-and-through guidewire. This form of ''blunt dissection'' was reported as useful by Lynch et al [74] in releasing 8 of 10 severely titled filters out of 257 retrievals.…”
Section: Balloon-displacement Techniquementioning
confidence: 99%