Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro

Abstract: Backside wear due to micromotion and poor conformity between the liner and its titanium alloy shell may contribute to the high rates of retroacetabular osteolysis and consequent aseptic loosening. The purpose of our study was to understand the wear process on the backside of polyethylene liners from two acetabular cup systems, whose locking mechanism is based on a press-fit cone in combination with a rough titanium conical inner surface on the fixation area. A direct comparison between in vitro wear simulator … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

1
7
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 10 publications
(8 citation statements)
references
References 34 publications
1
7
0
Order By: Relevance
“…The observers did not saw a clear change in the type and degree of wear as from the 5 to 10 million cycles analysis, but they did not had the previous results to influence their scoring. Regardless of the scores, the same wear modes and patterns were seen in the current in vitro study and in our previous retrieval study [11]. a previous study from pang et al [33] showed that visual backside damage scoring may not completely correspond with true volumetric wear, as no difference in damage scores were seen between xlpe and cpe liners, but there was three times more penetration in the cpe liners than in the xlpe liners.…”
Section: Discussionsupporting
confidence: 81%
See 2 more Smart Citations
“…The observers did not saw a clear change in the type and degree of wear as from the 5 to 10 million cycles analysis, but they did not had the previous results to influence their scoring. Regardless of the scores, the same wear modes and patterns were seen in the current in vitro study and in our previous retrieval study [11]. a previous study from pang et al [33] showed that visual backside damage scoring may not completely correspond with true volumetric wear, as no difference in damage scores were seen between xlpe and cpe liners, but there was three times more penetration in the cpe liners than in the xlpe liners.…”
Section: Discussionsupporting
confidence: 81%
“…a semi-quantitative method developed by Hood et al [15] and modified for hip implants was used to assess the damage on the backside of the liners. The detailed description of the modified method is given in the previously performed study [11] and will be briefly summarized in the following. On basis of its in situ orientation, the backside section was divided by a superior / inferior line and 7 different sections were determined ( Figure 1).…”
Section: Optical Backside Wear Analysismentioning
confidence: 99%
See 1 more Smart Citation
“…In addition, Braun et al verified that backside wear particles can occur in clinically established cup systems [15]. Furthermore, Reyna et al was able to detect backside wear caused by micromotion and poor conformity between cup and liner in an in vitro study [25]. As already mentioned, these particles could increase the potential of a resulting inflammatory reaction, which may result in osteolysis.…”
Section: Discussionmentioning
confidence: 91%
“…These results fit with other studies. Reyna et al compared the wear on the backside of PE (non‐XLPE and XLPE) liners used in a hip simulator test and retrieved liners using a semiquantitative scoring method and reported similar wear scores for both groups . Furthermore, Kurtz et al in 3D finite‐element models showed at least three orders of magnitude less volumetric wear rates from the backside of the liner compared to the volumetric wear rate estimates at the articulating bearing surface …”
Section: Discussionmentioning
confidence: 99%