“…After disconnecting them from the measuring signals a stable reference signal should be simultaneously connected to the inputs of the all MIs. Using the series of readings obtained from this set of MIs of the same kind -{Х 1 , Х 2 ,…, Х m } it is necessary to compute its arithmetic mean, which is considered to be a reference value: (1) Then, having determined errors of the MIs {Δ 1 , Δ 2 ,…,Δ m } and having compared values of these errors with maximum allowable limits for each MI there can be identified MIs whose error exceeds the allowable limits and, therefore, they should be removed from the process of measurements and be calibrated. An important requirement in the implementation of this method of comparison is to ensure the stability of the reference signal source while recording readings of all the MIs under test.…”