2013
DOI: 10.1063/1.4819607
|View full text |Cite
|
Sign up to set email alerts
|

Assessment of tungsten/rhenium thermocouples with metal-carbon eutectic fixed points up to 1500°C

Abstract: Four Type A thermocouples and two Type C thermocouples were calibrated at the Au fixed point and Co-C and Pd-C eutectic fixed points. The thermocouples were exposed to 1330 ºC for a total of 100 hours. The maximum drift due to the exposure was found to be 4.8 ºC. The fixed-point calibration EMF of these thermocouples deviated by less than 0.86% from the temperature specified by the standards ASTM E230-2003 for Type C and GOSTR 8.585-2001 for Type A. The length of one of Type A thermocouples A52 is longer than … Show more