Are Wireless Electronic Stethoscopes Useful for Respiratory Rate Monitoring During Intravenous Sedation?

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
3
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(3 citation statements)
references
References 16 publications
0
3
0
Order By: Relevance
“…Acoustic methods could achieve high accuracy in RR estimation. During intravenous sedation in dental procedures, there was no significant difference (p > 0.05) between the RR values derived by electrical stethoscopes and capnography (Morimoto et al 2018). For a total of 137 199 paired RR values measured by acoustic and capnography methods, the bias, standard deviation, and limits of agreement of respiration rate from acoustic monitoring versus capnography were 0.2 ± 2.3 bpm and −4.2 to 4.7 bpm, indicating good overall agreement of the two methods (Tanaka et al 2014).…”
Section: Acoustic Sensorsmentioning
confidence: 77%
See 1 more Smart Citation
“…Acoustic methods could achieve high accuracy in RR estimation. During intravenous sedation in dental procedures, there was no significant difference (p > 0.05) between the RR values derived by electrical stethoscopes and capnography (Morimoto et al 2018). For a total of 137 199 paired RR values measured by acoustic and capnography methods, the bias, standard deviation, and limits of agreement of respiration rate from acoustic monitoring versus capnography were 0.2 ± 2.3 bpm and −4.2 to 4.7 bpm, indicating good overall agreement of the two methods (Tanaka et al 2014).…”
Section: Acoustic Sensorsmentioning
confidence: 77%
“…Clinically, the capnometry-derived RR indicated limited ability to provide warning signs for a hypoxaemic event during the sedation procedure (Touw et al 2017). In some cases, new acoustic sensors showed better accuracy than capnometry in monitoring RR during anesthesia care (Tanaka et al 2014, Morimoto et al 2018. Therefore, more convenient capnometry devices, and higher accuracy in capnometry-derived RR during anesthesia, are still needed in further investigations.…”
Section: Humidity Sensors and Capnometrymentioning
confidence: 99%
“…OMS surgeons continue to practice outpatient anesthesia as first demonstrated by Dr. Morton in the Ether Dome. Newer monitoring devices such as the wireless precordial stethoscope(9) and bispectral index monitor (10) are available to detects signs of over sedation before clinical symptoms are present. The safety of intravenous anesthetics agents administered by OMS has been proven multiple times over…”