Application of glow discharge mass spectrometry to multielement ultra-trace determination in ultrahigh-purity copper and iron: a calibration approach achieving quantification and traceability

Abstract: A new approach was developed for quantitative calibration in GD-MS which can afford reliable and metrologically traceable results for many trace elements and was exemplified for pure copper and pure iron. It can be assumed that the technique can be further improved and applied to the analysis of other pure metals. Pressed copper and iron powder samples were used to calibrate the glow discharge mass spectrometry applied to the analysis of pure copper and iron. The new type of glow discharge mass spectrometer--t… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
47
0

Year Published

2007
2007
2019
2019

Publication Types

Select...
5
4
1

Relationship

1
9

Authors

Journals

citations
Cited by 62 publications
(49 citation statements)
references
References 20 publications
2
47
0
Order By: Relevance
“…The basic principle of the new methodology for the instrument calibration is to replace the current semi-quantitative approach by a quantitative one based on sets of doped samples with well-known impurity content whose values are directly traceable to the International System of Units (SI) and that provides measurement results with small uncertainty down to the ng · g −1 level ( Fig. 1) [9]. Such a methodology is the most accurate procedure to obtain small uncertainties using GD-MS. To date, calibration sets of high-purity zinc doped with eight different impurity fractions between 5 ng· g −1 and 10 µg · g −1 have been established.…”
Section: Glow Discharge Mass Spectrometrymentioning
confidence: 99%
“…The basic principle of the new methodology for the instrument calibration is to replace the current semi-quantitative approach by a quantitative one based on sets of doped samples with well-known impurity content whose values are directly traceable to the International System of Units (SI) and that provides measurement results with small uncertainty down to the ng · g −1 level ( Fig. 1) [9]. Such a methodology is the most accurate procedure to obtain small uncertainties using GD-MS. To date, calibration sets of high-purity zinc doped with eight different impurity fractions between 5 ng· g −1 and 10 µg · g −1 have been established.…”
Section: Glow Discharge Mass Spectrometrymentioning
confidence: 99%
“…With the aim of achieving highly accurate quantification with GDMS for rare elements and matrices, the calibration strategy with self-prepared synthetic standards was developed by Matschat et al 92 and then further improved, including evaluation of uncertainties, within the PhD thesis of Gusarova. 79 A similar concept has been applied to LA-ICP-MS quantitative analysis, as well.…”
Section: Calibration With Synthetic Standardsmentioning
confidence: 99%
“…GDs have been successfully coupled for analytical purposes to mass analysers such as quadrupoles [4] and double-focusing [5,6] spectrometers; however, the sequential nature of such mass analysers make them not ideal for depth profiling with high depth resolution. This sequential operation can be overcome using timeof-flight mass spectrometry (TOFMS) [7], capable of fast collection of a complete mass spectrum within one single extraction of the packet of ions (one extraction pulse of the analyser) with high precision and sensitivity.…”
Section: Introductionmentioning
confidence: 99%