2023
DOI: 10.3390/s23146263
|View full text |Cite
|
Sign up to set email alerts
|

Analysis of the Lifetime of Neural Implants Using In Vitro Test Structures

Abstract: The aim of this work was to measure the lifetime of neural implant test samples at two different temperatures, using a method that allows the precise measurement of the sample lifetime, further analysis with the use of Weibull statistics, and examination of the applicability of the Van’t Hoff rule. The correct estimation of the lifetime of neural implants is important to avoid preliminary failures, when used in humans. The novelty lies in the precise data due to the measurement approach, the application of the… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
0
0

Year Published

2024
2024
2024
2024

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(1 citation statement)
references
References 22 publications
0
0
0
Order By: Relevance
“…However, these changes were not very pronounced for the gold-coated needle electrodes during the immersion period, as also observed in other studies that investigated the longterm stability of gold in saline or phosphate solutions [43,44]. Determining the frequency of 1 kHz in vitro tests for implantable electrodes is important for evaluating their electrochemical behavior and tissue interaction, ensuring measurement stability, and achieving comparability of data across different studies and research [45,46]. The impedance at 1 kHz (Figure 14c(I,II)) remained unchanged after 21 days of immersion of gold-coated needles in saline solution at body temperature and after ultrasound treatments.…”
Section: Results Of In Vivo Testingsupporting
confidence: 54%
“…However, these changes were not very pronounced for the gold-coated needle electrodes during the immersion period, as also observed in other studies that investigated the longterm stability of gold in saline or phosphate solutions [43,44]. Determining the frequency of 1 kHz in vitro tests for implantable electrodes is important for evaluating their electrochemical behavior and tissue interaction, ensuring measurement stability, and achieving comparability of data across different studies and research [45,46]. The impedance at 1 kHz (Figure 14c(I,II)) remained unchanged after 21 days of immersion of gold-coated needles in saline solution at body temperature and after ultrasound treatments.…”
Section: Results Of In Vivo Testingsupporting
confidence: 54%