Coordination ionspray mass spectrometry (CIS-MS) is a useful tool in the detection and identification of cholesterol ester and phospholipid hydroperoxides and diacyl peroxides. Extensive studies of a series of cholesterol esters using CIS-MS revealed the following: (1) Cholesterol esters with equal number of double bonds as the internal standard showed a linear relative response in the mass spectrometer while compounds with non-equal numbers of double bonds gave a nonlinear relative response. (2) Complex adducts containing cholesterol ester, silver ion, AgF, AgBF 4 , and 2-propanoxide form when silver is in molar excess of cholesterol esters, reducing the [M ϩ Ag] ϩ signal. (3) In a mixture of cholesterol esters where silver is limiting, Ch22:6 and Ch20:4 bind to silver at the expense of Ch18:2 and have a higher signal in the mass spectrometer. (4) In a mixture of cholesterol esters where silver concentration is twofold greater than total cholesterol ester concentration, Ch22:6 and Ch20:4 form large complex adducts more frequently than Ch18:2 and have a lower signal in the mass spectrometer. (J Am Soc Mass Spectrom 2003, 14, 872-880)