2014
DOI: 10.1016/j.ejmp.2013.10.001
|View full text |Cite
|
Sign up to set email alerts
|

An audit of a hospital-based Doppler ultrasound quality control protocol using a commercial string Doppler phantom

Abstract: Abstract:Results from a four-year audit of a Doppler quality assurance (QA) program using a commercially available Doppler string phantom are presented. The suitability of the phantom was firstly determined and modifications were made to improve the reliability and quality of the measurements. QA of Doppler ultrasound equipment is very important as data obtained from these systems is used in patient management. It was found that if the braided-silk filament of the Doppler phantom was exchanged with an O-ring r… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
8
0

Year Published

2014
2014
2022
2022

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 9 publications
(8 citation statements)
references
References 12 publications
(14 reference statements)
0
8
0
Order By: Relevance
“…Complex flows are needed as a standard reference for the validation of innovative flow estimation algorithms. Recent audits concerning Ultrasound Quality Control protocols indicate that the current flow/velocity test objects available on the market are rather limited, expensive, and often fail to reproduce physiological conditions and physiological waveforms [15,16,17,18]. Manufacturers are not clear about tolerances and accuracy specifications [18] and the lack of well-defined protocols makes the Quality Control of scanners that measure flow and velocity challenging [15,16].…”
Section: Introductionmentioning
confidence: 99%
“…Complex flows are needed as a standard reference for the validation of innovative flow estimation algorithms. Recent audits concerning Ultrasound Quality Control protocols indicate that the current flow/velocity test objects available on the market are rather limited, expensive, and often fail to reproduce physiological conditions and physiological waveforms [15,16,17,18]. Manufacturers are not clear about tolerances and accuracy specifications [18] and the lack of well-defined protocols makes the Quality Control of scanners that measure flow and velocity challenging [15,16].…”
Section: Introductionmentioning
confidence: 99%
“…Of the main Doppler quality control tests maximum velocity accuracy has been found to be sensitive to element drop-out in a number of studies [38,44]. However, the question remains would the paper clip test provide this information before changes in maximum velocity accuracy have occurred.…”
Section: Discussionmentioning
confidence: 99%
“…This provided evidence of the importance of acceptance testing ultrasound systems, particularly in relation to the maximum velocity accuracy. In addition to maximum velocity accuracy, string phantoms have also been recommended by a number of investigators as a useful test device for other Doppler test parameters such as spectral broadening, lowest and highest detectable velocities, penetration depth, direction detection accuracy, range gate registration and sensitivity [7,39,43,44,47,48]. In 2006, Di Nallo et al presented a possible quality control procedure for use with a String Phantom.…”
Section: String Phantomsmentioning
confidence: 99%
See 2 more Smart Citations