An anomalous case of S-ICD malfunctioning: A big trouble or a soap bubble?

Abstract: Dvanáct hodin po implantaci subkutánního ICD přístroj aplikoval neadekvátní výboj pro detekci šumu. Kontrola ICD prokázala nedostatečné vnímání intrakardiálních signálů na primární vektor a významný zvukový artefakt na alternativním vektoru. Překvapivým zjištěním byla úplná obnova vnímání primárního vektoru po dvou dnech. Pravděpodobnou příčinou malfunkce byl vzduch uvězněný v podkožní kapse během implantace.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(3 citation statements)
references
References 4 publications
0
3
0
Order By: Relevance
“…implanting a replacement device in a subcutaneous pocket that previously housed a larger device may result in pocket air entrapment. Several authors 11,12,18,20–23 . following the instructions of the S‐ICD manufacturer 19 .…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…implanting a replacement device in a subcutaneous pocket that previously housed a larger device may result in pocket air entrapment. Several authors 11,12,18,20–23 . following the instructions of the S‐ICD manufacturer 19 .…”
Section: Discussionmentioning
confidence: 99%
“…Several authors. 11 , 12 , 18 , 20 , 21 , 22 , 23 following the instructions of the S‐ICD manufacturer. 19 suggest the subcutaneous injection of normal saline solution and the massage of skin to expel air from the subcutaneous space to prevent aerial entrapment during S‐ICD implantation.…”
Section: Discussionmentioning
confidence: 99%
“…These caused by the insulation of the sensing contact ring by the accumulation of air in the pocket. Recovery of sensing vector is expected in few days with the disappearance of the subcutaneous air [ 58 , 59 ].…”
Section: The S-icd Safetymentioning
confidence: 99%