Air kerma strength characterization of a GZP6 Cobalt-60 brachytherapy source

Abstract: The calculated and measured AKS values are in good agreement. Calculated-TPS and measured-TPS AKS values are also in agreement within the uncertainties related to our calculation, measurements and those certified by the GZP6 manufacturer. Considering the uncertainties, the TPS value for AKS is validated by our calculations and measurements, however, it is incorporated with a large uncertainty.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
7
0

Year Published

2012
2012
2018
2018

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 13 publications
(7 citation statements)
references
References 10 publications
0
7
0
Order By: Relevance
“…7 The GZP6 treatment planning system is able to produce dose distributions in the transverse and longitudinal planes. It calculates dose using the superposition dose calculation technique.…”
Section: Gzp6 Hdr Systemmentioning
confidence: 99%
See 2 more Smart Citations
“…7 The GZP6 treatment planning system is able to produce dose distributions in the transverse and longitudinal planes. It calculates dose using the superposition dose calculation technique.…”
Section: Gzp6 Hdr Systemmentioning
confidence: 99%
“…8 TG-43 recommended dosimetric characteristics of the sources in this system have been evaluated by several different investigators. [7][8][9] In this study channels 3 and 4 of GZP6 HDR system were utilized to deliver 1.5 Gy dose to 0.5 cm distance from the lateral surface of ovoids within 4.62 min. Each of the channels 3 and 4 contains one stationary active pellet (they are nearly identical in their source strength and geometry) and they can be used in the ovoids.…”
Section: Gzp6 Hdr Systemmentioning
confidence: 99%
See 1 more Smart Citation
“…1 were calculated using the DOSXYZnrc. 11,12 The voxel size used in the simulation was 0.1 mm × 0.1 mm × 0.1 mm, which corresponds to the x, y and z-axis. Under this calculation set up, the surface dose in this study was defined at the depth of 0.05 mm.…”
Section: Dose Calculation Of Phantoms Using the Dosxyznrcmentioning
confidence: 99%
“…using the MC method and GZP6 TPS. Toossi et al . estimated the air kerma strength of GZP6 60 Co source number 3 by Monte Carlo simulation and in‐air measurements.…”
Section: Introductionmentioning
confidence: 99%