Purpose:
To evaluate the performance of different nomograms and vault prediction formulas in predicting the optimal Implantable Collamer Lens (ICL; STAAR Surgical) size and vault.
Methods:
This retrospective study included 108 participants (214 eyes) who underwent ICL implantation. The efficacy of the manufacturer's nomogram, Parkhurst nomogram, NK formula, KS formula, Zhu formula, and ZZ formula was assessed by comparing the indicated ICL sizes to the postoperative vault measurements. Additionally, Bland-Altman plots and the Friedman test were used to assess the agreement and absolute error between the actual vault and predicted vault.
Results:
The proportions of the ideal ICL category recommended by the manufacturer's nomogram, Parkhurst nomogram, NK formula, KS formula, Zhu formula, and ZZ formula were 50.5%, 45.3%, 46.7%, 42.5%, 50.0%, and 28.5%, respectively. The mean differences between the actual vault and the predicted vault using the NK, KS, Zhu, and ZZ formulas were 144.1 ± 261.1, −19.3 ± 179.6, 70.8 ± 284.2, and 182.6 ± 361.5 μm, respectively. The predicted ICL vault tended to overestimate the actual ICL vault, particularly when choosing a larger ICL size. The KS formula shows the smallest deviation in prediction error and is least affected by variation in ICL size.
Conclusions:
White-to-white distance from the Pentacam (Oculus Optikgeräte GmbH) coupled with the manufacturer's nomogram performed well for ICL size selection. Four vault prediction formulas tended to overestimate the actual ICL vault, particularly when selecting a larger ICL size. The KS formula appeared to have the least bias of the formulas. Simultaneously, vault prediction formulas need to be modified according to the ICL size.
[
J Refract Surg
. 2023;39(7):456–461.]