The Clinical Assays kit was used as the "reference" kit against which the other 5 kits were compared, as this kit had been in routine use for 15 months. These kits showed good correlation with the reference kit (p < 0.001). Kits -D were tested to see if the high circulating steroids in the second half of a clinically normal pregnancy influenced kit performance. In these kits no "digoxin" level above 0.38 nmol · liter" 1 was recorded, despite the high cortisol In another 49 hospitalised patients undergoing regular digoxin therapy, the greatest discrepancy between these 4 kits in measuring digoxin in a single serum was 1.15 nmol · liter" 1 .In the kits employing antibody-coated tubes, modification of procedure often led to improved sensitivity and therefore to better precision in the lower concentration range.
Untersuchungen von sechs kommerziellen I2s j-markierten Digoxin-KitsZusammenfassung: Es wurden folgende kommerziell erhältlichen