“…For Tc-99 m, the supplier and receiver measurements were 25.0 mCi and 24.4 mCi (no correction in time), respectively, considering a time difference between them of - 10 min and a typical standard uncertainty (k = 1) of 0.72 % [ 75 ]. Based on the connected approaches of DR and guard band, the consumer risk for the supplier, receiver and reconciled measurements were, respectively, 20.2 % (not conforming to the specification), 6.15 % (not conforming to the specification) and 4.64 % (24.8 mCi–24.9 mCi, considering a standard deviation, conforming to the specification), generating reliable specification limits of 24.6 mCi–25.2 mCi, Fig.…”