Accuracy of a Portable Blood Gas Analyzer Incorporating Optodes for Canine Blood

Abstract: The accuracy of a portable blood gas analyzer (OPTI 1) was evaluated using canine blood and aqueous control solutions. Sixty-four arterial blood samples were collected from 11 anesthetized dogs and were analyzed for pH, partial pressure of carbon dioxide (PCO2) partial pressure of oxygen (PO2), and bicarbonate concentration ([HCO3-]) values by the OPTI 1 and a conventional blood gas analyzer (GASTAT 3). The conventional analyzer was considered as a standard against which the OPTI 1 was evaluated. Comparison of… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

1
0

Authors

Journals

citations
Cited by 1 publication
(1 citation statement)
references
References 6 publications
0
1
0
Order By: Relevance
“…In the present study, clinical agreement was determined by defining the clinically allowable limits for each variable on Bland-Altman plots using the following criteria required to pass proficiency testing for quality control 2,[34][35][36] : 80% of the data points should be within ±0.04 for pH, ±5.00 mm Hg for PCO 2 , and ±15.51 mm Hg for PO 2 on a standard analyzer. 35,36 The clinically allowable limits for [HCO 3 À ] were set at ±1.50 mmol/L to account for the analyzer's accuracy. 31 3 | RESULTS…”
Section: Discussionmentioning
confidence: 99%
“…In the present study, clinical agreement was determined by defining the clinically allowable limits for each variable on Bland-Altman plots using the following criteria required to pass proficiency testing for quality control 2,[34][35][36] : 80% of the data points should be within ±0.04 for pH, ±5.00 mm Hg for PCO 2 , and ±15.51 mm Hg for PO 2 on a standard analyzer. 35,36 The clinically allowable limits for [HCO 3 À ] were set at ±1.50 mmol/L to account for the analyzer's accuracy. 31 3 | RESULTS…”
Section: Discussionmentioning
confidence: 99%