Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films

Abstract: A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
8
0

Year Published

2016
2016
2022
2022

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 20 publications
(8 citation statements)
references
References 12 publications
0
8
0
Order By: Relevance
“…Dosemap is a software package developed by GE Healthcare to provide a live analytic estimation of PSD during procedure. 6 The outer shape of the patient is based on two super-ellipses model, built from real CT scan and defined by an equation previously described. 10,11 The local dose on this virtual envelope is then calculated as the following:…”
Section: Dosemap Softwarementioning
confidence: 99%
See 2 more Smart Citations
“…Dosemap is a software package developed by GE Healthcare to provide a live analytic estimation of PSD during procedure. 6 The outer shape of the patient is based on two super-ellipses model, built from real CT scan and defined by an equation previously described. 10,11 The local dose on this virtual envelope is then calculated as the following:…”
Section: Dosemap Softwarementioning
confidence: 99%
“…However, the use of these films in daily practice remains expensive and not easy to use. In this way, Dosemap software has been developed to estimate the PSD received by the patient in real time during the procedure . This system estimate the local dose distributed on the skin of the patient using dosimetric and technical information collected throughout the procedure.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…PSD can also be calculated from indirect dose metrics, such as the KA ref or the P KA 8-10 . On-line tools have been developed to monitor PSD and give a skin dose map 90 for radiation risk management during interventions 11,12 . Direct measurements of PSD can be performed using Gafchromic® films XR-RV3 5, 6, 12-16 .…”
Section: Introductionmentioning
confidence: 99%
“…CareMonitor by Siemens and DoseWise by Philips are solutions which basically provide the accumulated peak air kerma in the current projection 3 . On the other hand, Dose Map from GE is an advanced two‐dimensional (2D) solution, 4–6 while Dose Tracking System (DTS) from Canon (Toshiba) represents a state of the art three‐dimensional (3D) solution for skin dose mapping 7–9 . The most important drawback of previously mentioned solutions is that they are vendor specific and cannot be used in fluoroscopy systems from other manufacturers.…”
Section: Introductionmentioning