Accuracy evaluation of Roche Accu‐Chek Performa blood glucose meters at low glucose concentrations: A nine‐year retrospective study
Zhipeng Zhao,
Runqing Li,
Xiuying Zhao
et al.
Abstract:ObjectiveTo evaluate the accuracy of Roche Accu‐Chek Performa glucose meters at a low glucose concentration of <5.55 mmol/L (100 mg/dL) over a 9‐year period.MethodsThe accuracy of the Roche Accu‐Chek Performa glucose meters at low glucose concentrations was evaluated using annual comparison data for 9 consecutive years from 2015 to 2023, according to the acceptability criteria specified in International Organization for Standardization (ISO) 15197:2013. Blood samples with low glucose concentrations of <5… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.