2010
DOI: 10.1097/scs.0b013e3181d7ad06
|View full text |Cite
|
Sign up to set email alerts
|

Absorbable Plate Strength Loss During Molding

Abstract: Bioabsorbable plating systems play an integral role in cranial vault remodeling. After experiencing a case of plate failure requiring emergent reexploration, we investigated the potential causes. We hypothesize that extended submersion in the molding bath during plate preparation might advance the rate of hydrolysis and compromise plate structural integrity. Using an absorbable poly-D/L-lactic acid plating system, we assessed the effect of extended submersion on plate strength and stiffness when loaded in a ca… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
2
0

Year Published

2012
2012
2024
2024

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 11 publications
(2 citation statements)
references
References 10 publications
0
2
0
Order By: Relevance
“…Poly-llactide was used in the next generation, providing sufficient healing time before resorption. As the clinical usage of PLLA broadened from the late 1980s, reports documented successful applications in craniofacial surgeries, but problems such as weakened strength because of extended submersion during molding 1 and long intervals until complete degradation 8 still required attention. Bergsma et al reported that, even after 5.7 years, PLLA implants had not been fully resorbed.…”
Section: Discussionmentioning
confidence: 99%
“…Poly-llactide was used in the next generation, providing sufficient healing time before resorption. As the clinical usage of PLLA broadened from the late 1980s, reports documented successful applications in craniofacial surgeries, but problems such as weakened strength because of extended submersion during molding 1 and long intervals until complete degradation 8 still required attention. Bergsma et al reported that, even after 5.7 years, PLLA implants had not been fully resorbed.…”
Section: Discussionmentioning
confidence: 99%
“…It has been demonstrated that too much bending during plate contouring significantly weakens the material, as does leaving a plate too long in the water bath used to heat the plate for bending in the first place. 9 With regard to personal experience, I use resorbable plates in more cases in my pediatric practice than I did 10 years ago, but I use fewer plates. The advantages of reduced risks of intracranial migration and growth interference remain.…”
mentioning