2018
DOI: 10.1109/tmtt.2018.2851975
|View full text |Cite
|
Sign up to set email alerts
|

A Transmission Line Model for the Evaluation of MRI RF-Induced Fields on Active Implantable Medical Devices

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
18
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
6

Relationship

4
2

Authors

Journals

citations
Cited by 34 publications
(33 citation statements)
references
References 21 publications
0
18
0
Order By: Relevance
“…For proximal electrodes with small surface area, such as the ones in the SNM system, a smaller TF magnitude was observed for the bare-end configuration as shown in Figure 5C. 55 The correlation coefficients between the TF for the complete AIMD and the abandoned F I G U R E 3 Illustrations of three AIMDs and abandoned leads with different proximal end treatments. The blue rectangle is used to mask the implantable pulse generator (IPG) and lead names leads were calculated and are presented in Table 1.…”
Section: Transfer Function Measurement Resultsmentioning
confidence: 98%
“…For proximal electrodes with small surface area, such as the ones in the SNM system, a smaller TF magnitude was observed for the bare-end configuration as shown in Figure 5C. 55 The correlation coefficients between the TF for the complete AIMD and the abandoned F I G U R E 3 Illustrations of three AIMDs and abandoned leads with different proximal end treatments. The blue rectangle is used to mask the implantable pulse generator (IPG) and lead names leads were calculated and are presented in Table 1.…”
Section: Transfer Function Measurement Resultsmentioning
confidence: 98%
“…The RF‐induced heating for an implantable system can be assessed using the transmission line theorem, where the tissue itself acts as the return ground. Therefore, for different lead structures, the return ground is unchanged.…”
Section: Resultsmentioning
confidence: 99%
“…To address these questions we applied in this paper a rigorous method to develop the equivalent medium for accurate transfer function assessment. The entire AIMD lead was modeled as a transmission line where the two ends were terminated with a tip and an implantable pulse generator (IPG) impedance . The method to determine the equivalent medium is described in detail.…”
Section: Introductionmentioning
confidence: 99%
“…6 A detailed theoretical analysis of the problem was conducted by calculating the specific absorption rate around the electrode and the rise in temperature using the bioheat equation. 7 Moreover, theoretical models in the form of transfer function [8][9][10][11][12] or distributed circuit element model 9,13 have been proposed to analyze the problem and predict the temperature rise.…”
Section: Introductionmentioning
confidence: 99%