Mouth-to-microphone (MTM) distance is important when measuring the sound of voice. However, determining the MTM distance for laryngoscope-mounted microphones during laryngoscopic examinations is cumbersome. We introduce a novel solution for such cases, using the depth of insertion of the laryngoscope into the mouth DI as a reference distance. We measured the average insertion depth, DI, in 60 adult women and 60 adult men for rigid laryngoscopes with 70° and 90° view. We found the DI for the 70°/90° laryngoscope to be 9.7 ± 0.9/9.4 ± 0.6 cm in men, 8.9 ± 0.9/8.7 ± 0.7 cm in women, and 9.3 ± 0.9/9.0 ± 0.7 cm in all adults. Using these values, we show that, for microphones fixed at 15–40 cm from the tip of the laryngoscope, the final MTM distances are between 5 and 35 cm from the lips, and the standard uncertainties of these distances are between 16% and 2.5%. Our solution allows laryngologists and laryngoscope manufacturers to set and estimate the MTM distance for any rigid laryngeal endoscope with a microphone attached with reasonable accuracy, avoiding the need to measure this distance in vivo in routine practice.