A scalable correlator for multichannel diffuse correlation spectroscopy

Abstract: Diffuse correlation spectroscopy (DCS) is a technique which enables powerful and robust non-invasive optical studies of tissue micro-circulation and vascular blood flow. The technique amounts to autocorrelation analysis of coherent photons after their migration through moving scatterers and subsequent collection by single-mode optical fibers. A primary cost driver of DCS instruments are the commercial hardware-based correlators, limiting the proliferation of multi-channel instruments for validation of perfusio… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
8
0

Year Published

2016
2016
2021
2021

Publication Types

Select...
4
2

Relationship

4
2

Authors

Journals

citations
Cited by 9 publications
(8 citation statements)
references
References 6 publications
0
8
0
Order By: Relevance
“…This relationship between flow speed and a characteristic of the autocorrelation is all that is needed for DCS at its most basic level 3,5 . The phantom exhibits the pattern that it needs to replicate in order to be useful for baselining in vivo DCS experiments and instrument testing.…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…This relationship between flow speed and a characteristic of the autocorrelation is all that is needed for DCS at its most basic level 3,5 . The phantom exhibits the pattern that it needs to replicate in order to be useful for baselining in vivo DCS experiments and instrument testing.…”
Section: Resultsmentioning
confidence: 99%
“…A solid phantom remains inert and unaltered between experiments, allowing for effective benchmarking of instruments against a standard. A variety of solid and gel phantoms have been developed, although they generally only have a singular, large tube carrying liquid through the phantom 5,6 . This is not an accurate model of human tissue, where the majority of movement is blood flowing through capillaries.…”
Section: Introductionmentioning
confidence: 99%
“…For cost-effective construction of the device, DCS employs an inexpensive laser diode (LD) with a long coherence length to cover a sample volume. Instead of the commercialized hardware correlator, the correlator is implemented using a custom PSoc time-tagger and USB logger [10]. DRS uses a white LED as a broadband source and a miniaturized spectrometer (Avantes) to acquire a reflectance spectrum (450 ~ 650 nm).…”
Section: Methodsmentioning
confidence: 99%
“…A block diagram of the PSOC operating code is shown in Figure 3. Further details on the construction, design and operation of the DCS instrument is provided in [10]. …”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation