About a 4-fold increase in radioactive concentration of Tc was obtained (cf. approximately 10-12 ml normal saline is required for the elution of Tc from the gel column). Generators containing up to 22.2 GBq (600 mCi) Mo in 6-7 g ZrMo gel column (35 x 13 mm) were prepared and a radioactive concentration of Tc up to 4 GBq x ml (110 mCi x ml) was obtained on the first day of use. The overall recovery of Tc was >90%, Mo breakthrough was 10 to 10% and the duration of concentration was 3-5 min. The chemical impurity in terms of Al, Mo and Zr was <10 ppm each. The same procedure for the concentration of pertechnetate was applied to generators with 12-15 g ZrMo gel beds to obtain a higher capacity Tc gel generator, with similar findings.