A non-invasive device to objectively measure tibial rotation: verification of the device

Abstract: The purpose of this study was the correlation of the results of a new measurement device for tibial rotation (Rotameter) in comparison with the measurements of a knee navigation system as standard method. In a biomechanical laboratory study, all soft tissues were removed from 20 human cadaveric knees leaving only the intact capsule and the bone. Specific tracers were bicortically fixed in the bone in order to measure tibial rotation using a knee navigation system. The knees were fixed to a custom-made inside-b… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
50
0

Year Published

2010
2010
2017
2017

Publication Types

Select...
4
4

Relationship

1
7

Authors

Journals

citations
Cited by 34 publications
(51 citation statements)
references
References 26 publications
(46 reference statements)
1
50
0
Order By: Relevance
“…Hips are extended and knees flexed at 30°. The Rotameter overestimates the total range of rotation at 5, 10, and 15 Nm by an average of 5, 10, and 25°, respectively [33]. ICCs for these different torques were superior to 0.67 [34].…”
Section: Static Rotational Knee Laxitymentioning
confidence: 82%
See 1 more Smart Citation
“…Hips are extended and knees flexed at 30°. The Rotameter overestimates the total range of rotation at 5, 10, and 15 Nm by an average of 5, 10, and 25°, respectively [33]. ICCs for these different torques were superior to 0.67 [34].…”
Section: Static Rotational Knee Laxitymentioning
confidence: 82%
“…No data is available on its reproducibility. -Rotameter [33]: Two prototypes of the Rotameter exist. In both versions, the subject is lying prone to reproduce the dial test position.…”
Section: Static Rotational Knee Laxitymentioning
confidence: 99%
“…Six different measurement devices were identified in the recent literature to assess static rotational knee laxity [2,4,23,29,32,38]. The most important finding of the present review was that testing procedures and in vivo measurement conditions were highly variable, yielding results with up to threefold differences.…”
Section: Discussionmentioning
confidence: 96%
“…Table 1) [4,23,32,38]. The other two allowed for measurements at different knee flexion angles [2,29].…”
Section: Discussionmentioning
confidence: 99%
“…Other devices are dedicated to rotational measurements, like electronic goniometers [13] or gyroscopes and accelerometers [14,15]. Opto-electronic or magnetic 3D positioning techniques can assess both rotations and translations [16][17][18]. The measurement accuracy and reliability of these instruments have been widely studied [12,14,[16][17][18][19][20][21][22].…”
Section: Introductionmentioning
confidence: 99%