A new certified reference material for the quality control of palladium, platinum and rhodium in road dust, BCR-723

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
27
1

Year Published

2003
2003
2021
2021

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 67 publications
(30 citation statements)
references
References 16 publications
0
27
1
Order By: Relevance
“…42 This catalyst-associated PGE certied reference material is widely used in analytical laboratories examining various environmental materials (road dust, soil, sediment, and airborne particulate matter). The PGE-Re concentration and 187 Os/…”
Section: Paper Analytical Methodsmentioning
confidence: 99%
“…42 This catalyst-associated PGE certied reference material is widely used in analytical laboratories examining various environmental materials (road dust, soil, sediment, and airborne particulate matter). The PGE-Re concentration and 187 Os/…”
Section: Paper Analytical Methodsmentioning
confidence: 99%
“…It is officially known as BCR-723 and is available from the Institute for Reference Materials and Measurements (IRMM), Geel, Beligum [20,21]. The certified means (AESD) are: Pt, 81.3 AE 4.6 ng g À1 ; Pd, 6.0 AE 2.2 ng g À1 ; and Rh 12.8 AE 1.6 ng g À1 [21].…”
Section: Introductionmentioning
confidence: 99%
“…Since the introduction of BCR-723 (CW-8), the only statistical exploration of the concentration data submitted for PGE certification was provided by the original studies of Zischka et al [20,21]. The objectives of this study are to present a new quantitative exploration of the original PGE data with robust procedures, examine particular claims made in the original certification documents, and compare PGEs in BCR-723 with those representative of various European environmental media.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…One reason for this is the lack of analytical methods for Pd in the trace and ultratrace range, and the presence of severe mass interferences in inductively coupled plasma (ICP-MS) [15]. Recently, an effective separation and preconcentration method for Pd and Au was developed [16] and validated [17,18]. After sample decomposition both elements are collected by reductive coprecipitation with mercury.…”
Section: Introductionmentioning
confidence: 99%