2019
DOI: 10.1177/1533033819877986
|View full text |Cite
|
Sign up to set email alerts
|

A Method to Determine the Coincidence of MRI-Guided Linac Radiation and Magnetic Isocenters

Abstract: To assure accurate treatment delivery on any image-guided radiotherapy system, the relative positions and walkout of the imaging and radiation isocenters must be periodically verified and kept within specified tolerances. In this work, we first validated the multiaxis ion chamber array as a tool for finding the radiation isocenter position of a magnetic resonance–guided linear accelerator. The treatment couch with the array on it was shifted in 0.2-mm increments and the reported beam center position was plotte… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
18
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 14 publications
(19 citation statements)
references
References 14 publications
1
18
0
Order By: Relevance
“…The magnitude of the longitudinal (Y) isocenter shift did not exceed 0.11 mm and exhibited no strong gantry‐position‐dependent behavior. The static gantry‐position‐dependent isocenter displacement of the ViewRay 3D imaging was previously investigated by Latifi et al 14 In their paper, the minima and maximum of the vertical isocenter displacements occurred at the same positions that we report, and the horizontal isocenter exhibited a similarly minimal dependence on gantry position.…”
Section: Discussionsupporting
confidence: 74%
See 1 more Smart Citation
“…The magnitude of the longitudinal (Y) isocenter shift did not exceed 0.11 mm and exhibited no strong gantry‐position‐dependent behavior. The static gantry‐position‐dependent isocenter displacement of the ViewRay 3D imaging was previously investigated by Latifi et al 14 In their paper, the minima and maximum of the vertical isocenter displacements occurred at the same positions that we report, and the horizontal isocenter exhibited a similarly minimal dependence on gantry position.…”
Section: Discussionsupporting
confidence: 74%
“…The 3D data were used to establish the position of the phantom so that the geometric distortion in cine images could be evaluated. Because the magnetic susceptibility of the gantry is not uniformly distributed, the gantry position affects the homogeneity of the static magnetic field and can therefore produce geometric distortions 2,14 . For this reason, the static data were taken at multiple gantry positions in order to discriminate between geometric distortions arising from different static gantry positions and those arising from gantry motion.…”
Section: Methodsmentioning
confidence: 99%
“…Also included in this group of MLC-related checks were the gantry star shot with radiochromic film and a complementary gantry isocentricity test with an ion chamber array. 6 The latter two are used to define the radiation isocenter position at the time of installation.…”
Section: B | Mlc Testsmentioning
confidence: 99%
“…It has been shown previously that the Y-position readout correlated tightly with the radiation beam center measurement. 6…”
Section: B | Mlc Testsmentioning
confidence: 99%
“…4 There have been procedures in the literature describing methods to perform this measurement, based on film, polymer gel, spatial localization of individual ion chambers, or MR-compatible ion chamber arrays. [5][6][7] These methods can be somewhat cumbersome and time-consuming either with setup or post-exposure processing or rely on reduced information to infer 3D isocenter coincidence.…”
Section: Introductionmentioning
confidence: 99%