Abstract: In plant research, positron emission tomography (PET) is sometimes employed for physiological studies of plants. However, due to the generally high cost of PET systems, their application in plant research centers can be challenging. Furthermore, the configurations of commercial PET systems might not always align well with the requirements of plant research. To address these issues, we have taken the initial steps towards a solution by developing a large field-of-view (LFOV) angled panel detector … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
4
0

Year Published

2024
2024
2024
2024

Publication Types

Select...
1

Relationship

1
0

Authors

Journals

citations
Cited by 1 publication
(4 citation statements)
references
References 30 publications
(46 reference statements)
0
4
0
Order By: Relevance
“…This system utilizes eight panel detectors arranged in a ring to form a full-ring PET system with a detector ring diameter of 224 mm. Each panel detector comprises two gadolinium orthosilicate (GSO) blocks, an angled light guide, and four flat-panel photomultiplier tubes (FP-PMTs) [21]. The GSO pixel size for each block is 2.65 mm (radial) × 2.85 mm (axial) × 15.0 mm (depth), arranged in a 16 × 32 configuration for a GSO block.…”
Section: Concept Of the Plant Pet Systemmentioning
confidence: 99%
See 3 more Smart Citations
“…This system utilizes eight panel detectors arranged in a ring to form a full-ring PET system with a detector ring diameter of 224 mm. Each panel detector comprises two gadolinium orthosilicate (GSO) blocks, an angled light guide, and four flat-panel photomultiplier tubes (FP-PMTs) [21]. The GSO pixel size for each block is 2.65 mm (radial) × 2.85 mm (axial) × 15.0 mm (depth), arranged in a 16 × 32 configuration for a GSO block.…”
Section: Concept Of the Plant Pet Systemmentioning
confidence: 99%
“…Additionally, the hexadecagonal configuration simplifies the sorting and reconstruction software for the PET system, resembling the circular detector ring. Details of the design and performance of one panel detector to check the suitability for the development of a PET system were reported previously [21].…”
Section: Concept Of the Plant Pet Systemmentioning
confidence: 99%
See 2 more Smart Citations