A GATE Simulation Study of the Siemens Biograph DUO PET/CT System

Abstract: This is a GATE-simulation study of the Siemens Biograph DUO PET/CT system. It reports effects of changes in the thickness of the employed Lutetium Oxyorthosilicate (LSO) detectors. The PET/CT, a human body phantom and a cylindrical F-18 FDG source were simulated. Validation measurements were conducted. The results indicate that LSO thickness increase degrades spatial resolution, improves relative energy resolution from 9.0% to 11.3% and increases signal-to-noise-ratio from 0.81 to 1.17. Thicker LSO crystals pr… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(5 citation statements)
references
References 38 publications
(80 reference statements)
0
5
0
Order By: Relevance
“…The spectrum was convolved with the conventional LYSO Gaussian energy resolution at 511 keV (Surti et al 2007). 1) for GSO and LYSO PET, and extracted from Nikolopoulos et al (2013) for the LSO-PET. Blue line: 176 Lu-511 keV pile up component simulated with Gate; in B note the good reproduction of the observed increase before the hardware 850 keV cut-off.…”
Section: Long Energy Resolution Tail: Intrinsic Crystal Activity Comp...mentioning
confidence: 99%
See 1 more Smart Citation
“…The spectrum was convolved with the conventional LYSO Gaussian energy resolution at 511 keV (Surti et al 2007). 1) for GSO and LYSO PET, and extracted from Nikolopoulos et al (2013) for the LSO-PET. Blue line: 176 Lu-511 keV pile up component simulated with Gate; in B note the good reproduction of the observed increase before the hardware 850 keV cut-off.…”
Section: Long Energy Resolution Tail: Intrinsic Crystal Activity Comp...mentioning
confidence: 99%
“…Orange lines: energy spectra obtained by convolution of the gamma ray energies (171 and 245 keV for 111 In, and 511 keV for 18 F) with the energy resolution models (equations ( 7)-( 10), not shown for the LSO-PET). For GSO-PET and LYSO-PET specifications see material and method; for LSO-PET specifications see Nikolopoulos et al (2013).The higher tails amplitude for the LSO-PET is not linked to the crystal type, but is likely linked to the block design used in place of a continuous light guide for the GSO and LYSO PETs.…”
Section: Long Energy Resolution Tail: Intrinsic Crystal Activity Comp...mentioning
confidence: 99%
“…HR+ PET scanner and perform simulations based on voxelized phantom description [9]. Dimitrios Nikolopoulos et al [10] performed a GATE simulation for the Siemens Biograph DUO PET system. A human body phantom was simulated and the corresponding outputs were compared with the actual measurement.…”
Section: Jinst 12 T02001mentioning
confidence: 99%
“…For coincidence sorter policy, maximum energy is selected and the Coincidence time window is set as 12 ns, according to manufacturer setup. After these processing steps, the coincidence events were taken and stored in a ROOT file [10]. The digitizer chain allows us to tune the simulation to the very detailed parameters of the real PET setup.…”
Section: Signal Processingmentioning
confidence: 99%
“…Several clinical PET systems from leading manufacturers-Philips Healthcare, Siemens Healthineers, and General Electric Healthcare (GE) -have been modelled using GATE/Geant4, including Philips Allegro (Lamare et al 2006), Gemini TF (Trindade et al 2012), and Vereos (Salvadori et al 2020); Siemens Biograph 6 (Gonias et al 2007), DUO (Nikolopoulos et al 2013), mMR (Aklan et al 2015) and Vision (Zein et al 2020); and GE Discovery LS (Schmidtlein et al 2005), RX (Geramifar et al 2011), and MI (Kalaitzidis et al 2022, Tiwari et al 2022, Merlet et al 2024. A comprehensive review of GATE-modelled PET systems is provided in Sarrut et al (2021).…”
Section: Introductionmentioning
confidence: 99%