A Family of New Generation Miniaturized Impedance Analyzers for Technical Object Diagnostics

Abstract: The paper presents the family of three analyzers allowing to measure impedance in the range of 10 Ω<|Z x |<10 GΩ in a wide frequency range from 10 mHz up to 100 kHz. The most important features of the analyzer family are: miniaturization, low power consumption, low production cost, telemetric controlling and the use of an impedance measurement method based on digital signal processing (DSP). The miniaturization and other above-mentioned features of the analyzers were obtained thanks to the use of the newest ge… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
19
0

Year Published

2014
2014
2024
2024

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 31 publications
(20 citation statements)
references
References 7 publications
0
19
0
Order By: Relevance
“…Analog Devices propose to eliminate this error by the calibration [1,16]. Another proposal of how to deal with this problem is presented in a measurement circuit developed by Hoja and Lentka who used two AD5933 ICs in a new configuration different from the one recommended in the manufacturer's application note [12,13], where a simultaneous measurement of the voltage and current eliminates a need of the calibration cycle and removes the systematic error. We tried to analyze the factors on which this error depends and to compensate it mathematically.…”
Section: Systematic Admittance Argument Errormentioning
confidence: 99%
See 2 more Smart Citations
“…Analog Devices propose to eliminate this error by the calibration [1,16]. Another proposal of how to deal with this problem is presented in a measurement circuit developed by Hoja and Lentka who used two AD5933 ICs in a new configuration different from the one recommended in the manufacturer's application note [12,13], where a simultaneous measurement of the voltage and current eliminates a need of the calibration cycle and removes the systematic error. We tried to analyze the factors on which this error depends and to compensate it mathematically.…”
Section: Systematic Admittance Argument Errormentioning
confidence: 99%
“…An application of the AD5933 IC in biological diagnostics research has been reported by other authors. The AD5933 IC was used in the cell culture growth monitoring [2,3], single cell measurement [4], blood coagulation detection [5], biosensor applications [6], general bioimpedance measurements [7][8][9][10][11], and was also used in technical object monitoring [12,13]. It should be noted however, that in order to obtain the complete information about electrical properties of a measured object and to use a convenient method of the impedance spectra analysis (equivalent circuit modelling) the impedance has to be measured in a wide range of frequencies [14].…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Impedance analyser designs use either custom circuits [15] or application-specific integrated circuits (ICs), such as AD5833 [16]. In our case, since we need to study the effect of ADC resolution, we designed a simple specific circuit with monolithic ADC and DAC ICs.…”
Section: Design Of Test Circuitmentioning
confidence: 99%
“…The applications of impedance measurements in electrochemistry and the related field of corrosion monitoring [26,[41][42][43] require a particularly wide range of operating frequency, which is achieved by the same approach-microprocessor-controlled division of clock frequency [41,42].…”
Section: Corrosion and Electrochemistrymentioning