2022
DOI: 10.1039/d2an00743f
|View full text |Cite
|
Sign up to set email alerts
|

A direct mass spectrometry method for cannabinoid quantitation in urine and oral fluid utilizing reactive paper spray ionization

Abstract: A direct mass spectrometry method utilizing reactive paper spray ionization was developed for sensitive cannabinoid quantitation in biofluid matrices. The ca. 2-minute sample measurements used on-paper derivatization to significantly increase...

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 11 publications
(6 citation statements)
references
References 26 publications
0
6
0
Order By: Relevance
“…32 Qualifier ions were used to confirm albumin and creatinine results across spiked and real urine samples using a qualifier-to-quantifier ion ratio threshold of ±20%. Typical PS-MS experiments [33][34][35][36] require only 1 min of analysis to achieve satisfactory results (e.g., creatinine and creatinine-d 3 in Fig. 3).…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…32 Qualifier ions were used to confirm albumin and creatinine results across spiked and real urine samples using a qualifier-to-quantifier ion ratio threshold of ±20%. Typical PS-MS experiments [33][34][35][36] require only 1 min of analysis to achieve satisfactory results (e.g., creatinine and creatinine-d 3 in Fig. 3).…”
Section: Resultsmentioning
confidence: 99%
“…Typical PS-MS experiments 33–36 require only 1 min of analysis to achieve satisfactory results ( e.g. , creatinine and creatinine- d 3 in Fig.…”
Section: Resultsmentioning
confidence: 99%
“…Similarly, Borden et al used a reactive PS-MS method for the detection of cannabinoids in urine and saliva. 39 An on-paper derivatization method was used to increase the sensitivity of the technique, resulting in a limit of detection (LOD) of 0.8 ng/m for THC in saliva and 1.3 ng/ml for carboxy-THC in urine, the primary metabolite of THC. Finally, Vejar-Vivar et al…”
Section: Kim Et Al Used Sfpesi For the Rapid Detection Of Cocaine Met...mentioning
confidence: 99%
“…extraction). 11,[53][54][55] These and other strategies will be a focus for improving sensitivity and LODs for our methods in the future.…”
Section: Detection Of Arvs In Urine Samplesmentioning
confidence: 99%