2011
DOI: 10.4103/0019-5049.90637
|View full text |Cite
|
Sign up to set email alerts
|

A case of breakage of LMA Supreme TM on its first use

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
6
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(6 citation statements)
references
References 0 publications
0
6
0
Order By: Relevance
“…Simon BP have also reported intra-operative breakage of the LMA Supreme on its first use. The exact cause could not determined in their case [5]. After thorough evaluation in this case, we concluded that since the surgical site was close to the airway and the LMA was fixed in an erect position and was completely covered under surgical drapes, LMA breakage could have been caused by in advertent pressure exerted by the surgeons standing very close to the head end during axillary dissection, and the assistant surgeon's hand jerked suddenly due to malpositioning of the surgical retractor.…”
Section: Discussionmentioning
confidence: 87%
“…Simon BP have also reported intra-operative breakage of the LMA Supreme on its first use. The exact cause could not determined in their case [5]. After thorough evaluation in this case, we concluded that since the surgical site was close to the airway and the LMA was fixed in an erect position and was completely covered under surgical drapes, LMA breakage could have been caused by in advertent pressure exerted by the surgeons standing very close to the head end during axillary dissection, and the assistant surgeon's hand jerked suddenly due to malpositioning of the surgical retractor.…”
Section: Discussionmentioning
confidence: 87%
“…To our knowledge, this is only the second reported case of such a failure in a single‐use laryngeal mask , although there have been several case reports of breakages of re‐usable devices . Although our case resulted in no harm to the patient, there was potential for disaster.…”
mentioning
confidence: 71%
“…To our knowledge, this is only the second reported case of such a failure in a single-use laryngeal mask [1], although there have been several case reports of breakages of re-usable devices [2][3][4][5]. Although Figure 2 Yankauer sucker over a paediatric bougie to assist videolayngoscopic intubation.…”
Section: Kg Tuttlingen Germanymentioning
confidence: 99%
See 2 more Smart Citations