2018
DOI: 10.1088/1361-6439/aaceb5
|View full text |Cite
|
Sign up to set email alerts
|

A bimodal endoscopic imager in a glass package

Abstract: We present the first endoscopic imaging probe that combines full-field white light video microscopy and 3D optical coherence tomography (OCT) in a single miniaturized instrument. Designed as an instrument for the working channels of conventional endoscopes, the presented system has dimensions of mm3 (l  ×  w  ×  h) and is an important step towards practical multi-modal optical biopsy systems for clinical use. The video modality has a sensor-limited resolution and a total field of view of 10 μm and 1.17 mm in … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
10
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
6

Relationship

4
2

Authors

Journals

citations
Cited by 10 publications
(10 citation statements)
references
References 17 publications
0
10
0
Order By: Relevance
“…By continuously scaling the amplitude of the driving signals, a spiral scan pattern is obtained. To amplify the piezo displacement, the actuation frequency has to be close to the resonance frequency of the fiber . Figure B,D present a schematic depiction and image of the endoscopic probe head, respectively.…”
Section: Methodsmentioning
confidence: 99%
“…By continuously scaling the amplitude of the driving signals, a spiral scan pattern is obtained. To amplify the piezo displacement, the actuation frequency has to be close to the resonance frequency of the fiber . Figure B,D present a schematic depiction and image of the endoscopic probe head, respectively.…”
Section: Methodsmentioning
confidence: 99%
“…The lateral resolution of both modalities was measured using the knife-edge method and a resolution of 43.5 and 9.7 μm was determined for the OCT and WLI beam path, respectively. 23 Both channels are encapsulated in a glass package with outer dimensions of 15 × 2.4 × 3.3 mm 3 . The glass housing was manufactured by LightFab GmbH using a microstructuring process providing a precision of AE5 μm.…”
Section: Methodsmentioning
confidence: 99%
“…1(c). 23 For the acquisition of OCTA images, the piezoelectric tube was driven at an actuation frequency of 714 Hz at maximum voltage amplitude of 42.8 V to perform a spiral scan pattern with the fiber, and the circular FOV of the image was measured to be 1.146 mm in diameter. Before each measurement, a calibration of the piezo scanner was carried out to choose the best parameters for an optimum circular scan pattern.…”
Section: Methodsmentioning
confidence: 99%
“…By modulating the drive voltage in two pairs of electrodes, the desired scanning pattern from the fiber tip can be achieved. A large number of endoscopes are fabricated using this technique [84,91,[108][109][110].…”
Section: Piezoelectric Actuatorsmentioning
confidence: 99%
“…The angular motion is fast with respect to the radial motion. The scanning pattern is smooth and is used in fiber-optic endoscopic scanners [3,110]. A higher frame rate can be achieved using a larger amplitude compared to a raster scanner.…”
Section: Spiral Scanningmentioning
confidence: 99%