A 3-year review of MRI safety incidents within a UK independent sector provider of diagnostic services

Abstract: A review of MRI safety incidents conducted over a 3-year period for a large independent sector diagnostic imaging provider in the UK. The review took a systematic approach using reports logged on an internal incident reporting system that were then categorised and analysed for themes and trends. Notable cases and actions taken are also described from within the period. MRI safety-related events made up 7.5% of the total number of incident reports submitted and 15.5% of all MRI-related reports. The MR safety-re… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 8 publications
(8 citation statements)
references
References 20 publications
(45 reference statements)
0
7
0
Order By: Relevance
“…The risk of thermal injuries happening is primarily controlled through the design of the scanning equipment and whilst performing the examination. 6 Use of normal mode specific absorption rate (SAR) level during scanning is advocated for routine use which limits the whole body SAR to 2 Watts/kg, which in turn limits the rise in whole body temperature to no more than 0.5°C. 7 The indirect effects of heating are effectively managed through safe working practices, 6 such as safe positioning (pads), preparation (clothing) and other adjuncts to encourage natural cooling of the body (bore fan, controlled ambient temperature).…”
Section: Investigationsmentioning
confidence: 99%
“…The risk of thermal injuries happening is primarily controlled through the design of the scanning equipment and whilst performing the examination. 6 Use of normal mode specific absorption rate (SAR) level during scanning is advocated for routine use which limits the whole body SAR to 2 Watts/kg, which in turn limits the rise in whole body temperature to no more than 0.5°C. 7 The indirect effects of heating are effectively managed through safe working practices, 6 such as safe positioning (pads), preparation (clothing) and other adjuncts to encourage natural cooling of the body (bore fan, controlled ambient temperature).…”
Section: Investigationsmentioning
confidence: 99%
“… 18 Based on the analyzed MRI safety reports, referral of contraindicated devices (32% of reports) and a failure to verify safety (21.5%) were identified as the two main trends. 19 A Swedish study found that incidents involving MRIs are commonly underreported. 20 MRI accidents can be minimized by implementing strict guidelines, effective management, and establishing MRI safety policies.…”
Section: Introductionmentioning
confidence: 99%
“…Westbrook and Talbot measured radiographers' knowledge about MR, and they found that MR safety knowledge in particular was very low [11]. Hudson et al also pointed out that the low level of MR safety knowledge of radiographers could cause incidents [12].…”