Study on the Performance of an Alpha Energy Spectrum Radon Measuring Instrument Based on CdZnTe Detector

Abstract: To realize a fast and reliable approach for continuous radon measurements, a CdZnTe (CZT) detector based on the electrostatic collection method was developed. The experimental results show that when the external temperature varies from 5 °C to 30 °C, the maximum drifts of the characteristic peak positions of 218Po and 214Po are only 8 and 6. Furthermore, the measurement error associated with the constant radon concentration is less than 5.5%. As the temperature increases or decreases, the measurement error bec… Show more