2018
DOI: 10.1111/pace.13513
|View full text |Cite
|
Sign up to set email alerts
|

AV hysteresis causing initiation of recurrent atrial arrhythmias

Abstract: A 73‐year‐old male with dual‐chamber implantable cardioverter defibrillator (Teligen, Boston Scientific, Marlborough, MA, USA) had multiple episodes of automatic mode switch (AMS) during clinical follow‐up. Over 50% of these demonstrated a similar pattern of initiation. AV Search+ in combination with sensor rate pacing can cause short‐coupled atrial paced intervals that can be proarrhythmic. After programming changes were made AMS burden has significantly decreased.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2022
2022
2022
2022

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
(1 citation statement)
references
References 4 publications
(7 reference statements)
0
1
0
Order By: Relevance
“…Similar incidences with RYTHMIQ™ were previously reported to be proarrhythmic, though these were as a result of short-long-short coupling intervals. 7 , 8 …”
Section: Discussionmentioning
confidence: 99%
“…Similar incidences with RYTHMIQ™ were previously reported to be proarrhythmic, though these were as a result of short-long-short coupling intervals. 7 , 8 …”
Section: Discussionmentioning