2014
DOI: 10.1016/j.medengphy.2014.05.013
|View full text |Cite
|
Sign up to set email alerts
|

Optimal needle design for minimal insertion force and bevel length

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
36
0
1

Year Published

2016
2016
2024
2024

Publication Types

Select...
8

Relationship

2
6

Authors

Journals

citations
Cited by 31 publications
(37 citation statements)
references
References 15 publications
0
36
0
1
Order By: Relevance
“…For the baseline NLP, a commercial lancet needle (Becton-Dickinson) is selected. By using the reverse engineering method [7], the baseline lancet needle, with ξ 15°, ϕ 20°, and β 60°, is also ground. Figure 4 shows the final fabricated results of FLN-B, FLN-F, and regular NLP needles.…”
Section: Fln-b and Fln-f Needles Fabricationmentioning
confidence: 99%
See 1 more Smart Citation
“…For the baseline NLP, a commercial lancet needle (Becton-Dickinson) is selected. By using the reverse engineering method [7], the baseline lancet needle, with ξ 15°, ϕ 20°, and β 60°, is also ground. Figure 4 shows the final fabricated results of FLN-B, FLN-F, and regular NLP needles.…”
Section: Fln-b and Fln-f Needles Fabricationmentioning
confidence: 99%
“…Mathematical models to quantify the needle cutting edge inclination (λ) and rake (α) angles have been developed, and results demonstrated that two lancets at the needle tip can increase the λ and lower the needle insertion force [4]. Based on the model to estimate the initial peak needle insertion force [5,6], optimal lancet needle geometry can be achieved, which has large lancets on the tip and could reduce the needle insertion force by 14% [7]. Along the needle tip cutting edge, two parameters, the λ and α, have demonstrated to be critical for the needle insertion force [5,8].…”
Section: Introductionmentioning
confidence: 99%
“…The rake and inclination angle are two key parameters that define the geometry and orientation of the cutting edge [53] and can also affect the needle insertion force [54]. Using a mathematical model to predict the needle insertion force [55], an optimization algorithm can be applied to vary the parameters that define the tip geometry of a needle. Results demonstrate that needles with a large bevel surface have the lowest insertion force [55].…”
Section: Tissue Acquistionmentioning
confidence: 99%
“…In Figure 2, which presents the setup of the experiment for measuring the acceleration measurement of the lancing device, hop (1) clamps a lancing device (2), and an accelerometer (3) is used to measure the acceleration of the lancet immediately after it strikes the lancing device. The acceleration and velocity data are transmitted to an acquisition device (4), which is connected to a computer (5). Subsequent analysis is performed using Pico Scope 6 (2013, PICO Technology, Eaton Socon, England) and Matlab software packages (Version 7, The MathWorks, Natick, MA, USA, 2014).…”
Section: Lancing Device Principlementioning
confidence: 99%
“…(1) Change the shape and bevel angle of the lancet tip to reduce the size of the wound that is produced by piercing [5,6]. (2) Coat the lancet surface with a special material to reduce the friction between the skin and the lancet and thereby to reduce pain [7].…”
Section: Introductionmentioning
confidence: 99%