Flow induced displacement forces in iliac limb stent grafts are significant and are influenced by distal stent graft diameter and the shape of the graft curvature. The displacement forces are particularly high at the large distal end of bell-bottom grafts. Wide iliac arteries treated with bell-bottom stent grafts may require more vigilant surveillance and improved stent graft fixation.