2016
DOI: 10.5935/1678-9741.20160036
|View full text |Cite
|
Sign up to set email alerts
|

Duration of systole and diastole for hydrodynamic testing of prosthetic heart valves: comparison between ISO 5840 standards and in vivo studies

Abstract: ObjectiveTo complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review.MethodsThe systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves.ResultsBased on the in vivo studies analyzed, the systolic durat… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
5
0

Year Published

2017
2017
2020
2020

Publication Types

Select...
4
2
1

Relationship

1
6

Authors

Journals

citations
Cited by 7 publications
(5 citation statements)
references
References 9 publications
0
5
0
Order By: Relevance
“…We used a waveform with an approximately 50% systolic duration to better match in vivo conditions. 27 Pressure transducers and flow meters were zeroed before trials, and the baseline valve was reinstalled between experiments to reduce drift.…”
Section: Left Heart Simulator Design and Validationmentioning
confidence: 99%
“…We used a waveform with an approximately 50% systolic duration to better match in vivo conditions. 27 Pressure transducers and flow meters were zeroed before trials, and the baseline valve was reinstalled between experiments to reduce drift.…”
Section: Left Heart Simulator Design and Validationmentioning
confidence: 99%
“…In healthy patients the systole percent is about 25–35% of the cardiac cycle. 21 In patient suffering from heart failure the systole percent can reach 73% of the cardiac cycle. 22 …”
Section: Discussionmentioning
confidence: 99%
“…The LVM was established for a heart rate of 60 bpm with a systolic duration of 400 ms , a stroke volume of 70 mL, and an end‐diastolic volume of 120 mL. The aortic pressure (AoP) was increased successively, until it reached the range of 80–120 mm Hg , as shown in Fig.…”
Section: Methodsmentioning