“…The relative standard deviations for eleven replicate determinations of 0.1 and 0.25 mg L -1 were 3.2 and 2.4%, respectively. A comparable analytical performance of the proposed method with other flowbased methods for the spectrophotometric determination of aluminium, using various synthetic dyes, such as chrome azurol S (CAS), 11,32,33 eriochrome cyanine R (ECR), 34,35 and quercetin 36 was obtained, although, the developed method used only natural dye as a chromogenic reagent. In addition, the developed method provided a higher sensitivity than the previous flowbased methods using other natural reagents such as Morinda citrifolia root extract, 24 and Terminalia catappa L. leaf extract.…”