We evaluated Clearline Influenza A/B/(H1N1)2009, a new multi-line immunochromatographic assay for rapid detection of antigens of influenza A (Flu A), B (Flu B), and A(H1N1)2009 viruses. Clearline detected Flu A, Flu B, and A(H1N1)2009 viruses with a detection limit of 4.6 × 103 to 7.5 × 104 pfu/assay. The sensitivity and specificity of detection of influenza virus by Clearline, using RT-PCR as reference standard, were determined for A(H1N1)2009, Flu A, and Flu B, in nasopharyngeal aspirate, nasopharyngeal swab, and self-blown nasal discharge specimens. Sensitivity for nasopharyngeal aspirate specimens was: A(H1N1)2009 = 97.3 %, Flu A = 94.5 %, and Flu B = 96.8 %, and specificity was Flu A = 99.1 % and Flu B = 100 %. Sensitivity for nasopharyngeal swab specimens was: A(H1N1)2009 = 91.9 %, Flu A = 92.8 %, and Flu B = 100 %, and specificity was Flu A = 98.2 % and Flu B = 100 %. Sensitivity for self-blown nasal discharge specimens was: A(H1N1)2009 = 75.7 %, Flu A = 86.5 %, and Flu B = 76.2 %, and specificity was Flu A = 98.4 % and Flu B = 100 %. Sensitivity and specificity of Clearline were sufficient for nasopharyngeal aspirate and swab specimens. For self-blown nasal discharge specimens, sensitivity was lower than for nasopharyngeal aspirates and nasopharyngeal swabs. The sensitivity of Clearline for A(H1N1)2009 was good even 6 h after the onset of symptoms. These findings suggest that Clearline may be useful for early clinical diagnosis of influenza.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.