Aim
To evaluate the use of hybrid closed‐loop glucose control with faster‐acting insulin aspart (Fiasp) in adults with type 1 diabetes (T1D).
Research Design and Methods
In a double‐blind, multinational, randomized, crossover study, 25 adults with T1D using insulin pump therapy (mean ± SD, age 38 ± 9 years, HbA1c 7.4% ± 0.8% [57 ± 8 mmol/mol]) underwent two 8‐week periods of unrestricted living comparing hybrid closed‐loop with Fiasp and hybrid closed‐loop with standard insulin aspart in random order. During both interventions the CamAPS FX closed‐loop system incorporating the Cambridge model predictive control algorithm was used.
Results
In an intention‐to‐treat analysis, the proportion of time sensor glucose was in the target range (3.9–10.0 mmol/L; primary endpoint) was not different between interventions (75% ± 8% vs. 75% ± 8% for hybrid closed‐loop with Fiasp vs. hybrid closed‐loop with standard insulin aspart; mean‐adjusted difference −0.6% [95% CI −1.8% to 0.7%]; p < .001 for non‐inferiority [non‐inferiority margin 5%]). The proportion of time with sensor glucose less than 3.9 mmol/L (median [IQR] 2.4% [1.2%–3.2%] vs. 2.9% [1.7%–4.0%]; p = .01) and less than 3.0 mmol/L (median [IQR] 0.4% [0.2%–0.7%] vs. 0.7% [0.2%–0.9%]; p = .03) was reduced with Fiasp versus standard insulin aspart. There was no difference in mean glucose (8.1 ± 0.8 vs. 8.0 ± 0.8 mmol/L; p = .13) or glucose variability (SD of sensor glucose 2.9 ± 0.5 vs. 2.9 ± 0.5 mmol/L; p = .90). Total daily insulin requirements did not differ (49 ± 15 vs. 49 ± 15 units/day; p = .45). No severe hypoglycaemia or ketoacidosis occurred.
Conclusions
The use of Fiasp in the CamAPS FX closed‐loop system may reduce hypoglycaemia without compromising glucose control compared with standard insulin aspart in adults with T1D.
AimTo evaluate the PAQ
® (CeQur SA, Horw, Switzerland), a wearable 3‐day insulin delivery device that provides set basal rates and bolus insulin on demand, in people with Type 2 diabetes.MethodAdults with Type 2 diabetes with HbA1c concentrations ≥53 and ≤97 mmol/mol (7.0 and 11.0%) while treated with ≥2 insulin injections/day were enrolled in two single‐arm studies comprising three periods: a baseline (insulin injections), a transition and a PAQ treatment period (12 weeks). Endpoints included HbA1c, seven‐point self‐monitored blood glucose, total daily dose of insulin and body weight. Safety was assessed according to examination, hypoglycaemic episodes and adverse device effects.ResultsA total of 28 adults were enrolled (age 63 ± 7 years, 86% men, BMI 32.3 ± 4.3kg/m2, Type 2 diabetes duration 17 ± 8 years, HbA1c 70 ± 12 mmol/mol (8.6 ± 1.1%), total daily insulin dose 58.7 ± 20.7 U), of whom 24 completed the studies. When transitioned to PAQ, 75% of participants continued on the first basal rate selected. After 12 weeks of PAQ wear, significant improvements from baseline were seen [HbA1c –16 ± 9 mmol/mol (95% CI –20, –12) or –1.5 ± 0.9% (95% CI –1.8, –1.1) P<0.0001], and at all seven self‐monitored blood glucose readings time points (P ≤0.03). Total daily insulin dose increased by 12.1 ± 19.5 U (95% CI 3.9, 20.4; P=0.0058), the number of meal time boluses increased by 0.9 ± 1.5/day (95% CI 0.3, 1.5; P=0.0081) and body weight remained stable. Six participants had mild to moderate catheter site reactions and one mild skin irritation occurred. No participant experienced severe hypoglycaemia.ConclusionsAdults with Type 2 diabetes were safely transitioned from insulin injections to the PAQ and had significantly improved glycaemic control and treatment satisfaction with insulin therapy.(ClinicalTrials.gov identifiers: NCT02158078 & NCT02419859)
Objective: Many hybrid closed-loop (HCL) systems struggle to manage unusually high glucose levels as experienced with intercurrent illness or pre-menstrually. Manual correction boluses may be needed, increasing hypoglycemia risk with overcorrection. The Cambridge HCL system includes a user-initiated algorithm intensification mode (“Boost”), activation of which increases automated insulin delivery by approximately 35%, while remaining glucose-responsive. In this analysis, we assessed the safety of “Boost” mode. Methods: We retrospectively analyzed data from closed-loop studies involving young children (1-7 years, n = 24), children and adolescents (10-17 years, n = 19), adults (≥24 years, n = 13), and older adults (≥60 years, n = 20) with type 1 diabetes. Outcomes were calculated per participant for days with ≥30 minutes of “Boost” use versus days with no “Boost” use. Participants with <10 “Boost” days were excluded. The main outcome was time spent in hypoglycemia <70 and <54 mg/dL. Results: Eight weeks of data for 76 participants were analyzed. There was no difference in time spent <70 and <54 mg/dL between “Boost” days and “non-Boost” days; mean difference: –0.10% (95% confidence interval [CI] –0.28 to 0.07; P = .249) time <70 mg/dL, and 0.03 (–0.04 to 0.09; P = .416) time < 54 mg/dL. Time in significant hyperglycemia >300 mg/dL was 1.39 percentage points (1.01 to 1.77; P < .001) higher on “Boost” days, with higher mean glucose and lower time in target range ( P < .001). Conclusions: Use of an algorithm intensification mode in HCL therapy is safe across all age groups with type 1 diabetes. The higher time in hyperglycemia observed on “Boost” days suggests that users are more likely to use algorithm intensification on days with extreme hyperglycemic excursions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.